Dental cone-beam computed tomography acquires projection data using a cone-shaped x-ray beam and reconstructs volumetric images for dentistry and related head-and-neck applications. Protocol selection should be determined by the clinical task, with attention to field of view, exposure settings, and patient characteristics.
Dental cone-beam computed tomography systems should not automatically be evaluated using methods intended for conventional multidetector computed tomography. Geometric alignment, collimation, spatial resolution, uniformity, noise, artifacts, and relevant radiation-output measures require equipment-appropriate methods.
American Association of Physicists in Medicine Task Group 261 provides a comprehensive framework for dental/maxillofacial cone-beam computed tomography quality control and discusses the role of medical physicists, acceptance testing, ongoing testing, and radiation protection. Regulatory and accreditation requirements vary by jurisdiction.
Source: American Association of Physicists in Medicine Task Group 261 (2024)